Automated 1D Helmholtz coil design for cell biology: Weak magnetic fields alter cytoskeleton dynamics

Abasalt Bahrami1,2    Leonardo Y. Tanaka2    Ricardo C. Massucatto2    Francisco R. M. Laurindo2    Clarice D. Aiello3 clarice@quantumbiology.tech 1Department of Electrical and Computer Engineering, University of California, Los Angeles, CA, 90095 US
2Vascular Biology Laboratory, Heart Institute (InCor), University of São Paulo School of Medicine, São Paulo, Brazil
3Quantum Biology Tech (QuBiT) Lab, Los Angeles, 90095 US
(June 27, 2024)
Abstract

For more than fifty years, scientists have been gathering evidence of the biological impacts of weak magnetic fields. However, the lack of systematics in experimental studies has hampered research progress on this subject. To systematically quantify magnetic field effects in cell biology, it is crucial to produce fields that can be automatically adjusted and that are stable throughout an experiment’s duration, usually operating inside an incubator. Here, we report on the design of a fully automated 1D Helmholtz coil setup that is internally water cooled, thus eliminating any confounding effects caused by temperature fluctuations. The coils also allow cells to be exposed to magnetic fields from multiple directions through automated controlled rotation. Preliminary data, acquired with the coils placed inside an incubator and on a rat vascular smooth muscle cell line, confirm previous reports that both microtubule and actin polymerization and dynamics are altered by weak magnetic fields.

preprint: APS/123-QED

I Introduction

The scientific community has long been intrigued by the effects of magnetic fields (MFs) on biology across the tree of life, as reported by numerous studies [1, 2, 3, 4, 5]. Despite extensive research spanning more than half a century, the mechanisms underlying the observed effects on cellular systems remain unclear and ambiguous. In some instances, the effects themselves are contentious or difficult to replicate. Even though cell biology also seem to respond to both strong fields and MF gradients, in this work we focus on the biological effects of uniform and weak MFs [6, 7, 8]. By weak, we mean fields whose intensity are up to approximately ten times that of the Earth’s MF, which is around 25–65 μT𝜇T\mu\text{T}italic_μ T.

Weak MFs impact a myriad of cellular processes, including the regulation of gene expression [9], ion channel functioning [10], and the production of reactive oxygen species [11, 12, 13, 14, 15]. MFs can also impact apoptosis in both cancer and normal cells [16, 17, 18, 19, 20, 21]. MFs influence metabolic pathways [22], inducing changes in cell growth rate [23, 24], affecting cellular morphology [25, 26, 27, 28], and promoting muscle cell development [29]. Finally, it is important to recognize that MF effects do not necessarily monotonically increase with the strength of the applied field: higher magnetic field intensities do not necessarily lead to more pronounced effects [6, 8].

As our environment becomes more polluted by electromagnetic fields from various sources, understanding the effects of being exposed to these small-magnitude fields is a timely endeavor. Moreover, such an electromagnetic ‘knob’ in biology could be, and has already been, rationally used for medical theranostics [30, 31], with magnetic therapies proposed to accelerate diabetic wound repair [32] and pain management [33], and to counter neurodegenerative diseases [34, 35] and cancer [36, 37, 38, 39].

To unambiguously establish and understand the biological effects of weak MFs, it is crucial to perform well-controlled and reproducible experiments. Some experiments on bioelectromagnetics use permanent magnets, facing challenges such as fixed field strengths and non-uniformity. In contrast, coil setups, such as those obeying a Helmholtz configuration, offer the advantage of controllably changing uniform MFs [40, 41, 42]. Nevertheless, most current designs limit the MF application to a single direction, requiring manual coil rotation for direction changes. Our new design of 1D Helmholtz coils for bioelectromagnetic experiments overcomes the constraints of traditional setups by introducing complete automation of the MF direction. Additionally, we incorporate an effective cooling mechanism to mitigate temperature rise effects on cells. Many currently employed coils in bioelectromagnetic experiments lack an active cooling system, posing challenges for conducting prolonged experiments under sustained MFs [43, 44, 45] and maintaining stable temperature conditions.

II Coil design

We designed a set of Helmholtz coils with water cooling. Unlike conventional coils that often, during operation, produce temperature fluctuations [46], this design efficiently prevents temperature rise by allowing for the regulation of the water flow rate (Suppl. Fig. SF1). This set of coils can be controllably rotated around its pivot using a servo motor. We could not detect any MFs generated by the servo motor within our measurement accuracy of (1.0 ±plus-or-minus\pm± 0.1) μ𝜇\muitalic_μT. By allowing the coil assembly to rotate, biological samples could be subjected to uniform 1D MFs at various orientations.

We chose a design capable of producing MF as strong as 15 mT, with this upper limit being determined by the size of the copper wire thickness 0.71 mm (22 gauge wire) chosen to wound the coils. Magnetic fields were numerically simulated using the Biot-Savart law. The axial and radial MFs produced by each coil in the Helmholtz configuration are mathematically given by [47]:

Basubscript𝐵𝑎\displaystyle B_{a}italic_B start_POSTSUBSCRIPT italic_a end_POSTSUBSCRIPT =B0πQ[E(k)1α2β2Q4α+K(k)],absentsubscript𝐵0𝜋𝑄delimited-[]𝐸𝑘1superscript𝛼2superscript𝛽2𝑄4𝛼𝐾𝑘\displaystyle=\frac{B_{0}}{\pi\sqrt{Q}}\left[E(k)\frac{1-\alpha^{2}-\beta^{2}}% {Q-4\alpha}+K(k)\right],= divide start_ARG italic_B start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT end_ARG start_ARG italic_π square-root start_ARG italic_Q end_ARG end_ARG [ italic_E ( italic_k ) divide start_ARG 1 - italic_α start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT - italic_β start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT end_ARG start_ARG italic_Q - 4 italic_α end_ARG + italic_K ( italic_k ) ] , (1)
Brsubscript𝐵𝑟\displaystyle B_{r}italic_B start_POSTSUBSCRIPT italic_r end_POSTSUBSCRIPT =B0γπQ[E(k)1+α2+β2Q4αK(k)],absentsubscript𝐵0𝛾𝜋𝑄delimited-[]𝐸𝑘1superscript𝛼2superscript𝛽2𝑄4𝛼𝐾𝑘\displaystyle=\frac{B_{0}\gamma}{\pi\sqrt{Q}}\left[E(k)\frac{1+\alpha^{2}+% \beta^{2}}{Q-4\alpha}-K(k)\right],= divide start_ARG italic_B start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT italic_γ end_ARG start_ARG italic_π square-root start_ARG italic_Q end_ARG end_ARG [ italic_E ( italic_k ) divide start_ARG 1 + italic_α start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT + italic_β start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT end_ARG start_ARG italic_Q - 4 italic_α end_ARG - italic_K ( italic_k ) ] , (2)

where

Basubscript𝐵𝑎\displaystyle B_{a}italic_B start_POSTSUBSCRIPT italic_a end_POSTSUBSCRIPT is the field component aligned with the axis [G],
Brsubscript𝐵𝑟\displaystyle B_{r}italic_B start_POSTSUBSCRIPT italic_r end_POSTSUBSCRIPT is the field component in a radial direction [G],
a0subscript𝑎0\displaystyle a_{0}italic_a start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT is the loop radius [m],
μ0subscript𝜇0\displaystyle\mu_{0}italic_μ start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT is the permeability constant [4π×107NA2],\displaystyle\text{ is the permeability constant [}4\pi\times 10^{-7}\,\text{N% }\cdot\text{A}^{-2}],is the permeability constant [ 4 italic_π × 10 start_POSTSUPERSCRIPT - 7 end_POSTSUPERSCRIPT N ⋅ A start_POSTSUPERSCRIPT - 2 end_POSTSUPERSCRIPT ] ,
x𝑥\displaystyle xitalic_x is the axial distance from the center of the coil [m],
r𝑟\displaystyle ritalic_r is the radial distance from the axis of the coil [m],
α𝛼\displaystyle\alphaitalic_α =r/a0,β=x/a0,γ=x/r,formulae-sequenceabsent𝑟subscript𝑎0formulae-sequence𝛽𝑥subscript𝑎0𝛾𝑥𝑟\displaystyle=r/a_{0},\quad\beta=x/a_{0},\quad\gamma=x/r,= italic_r / italic_a start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT , italic_β = italic_x / italic_a start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT , italic_γ = italic_x / italic_r ,
Q𝑄\displaystyle Qitalic_Q =(1+α)2+β2,absentsuperscript1𝛼2superscript𝛽2\displaystyle=(1+\alpha)^{2}+\beta^{2},= ( 1 + italic_α ) start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT + italic_β start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT ,
k𝑘\displaystyle kitalic_k =4αQ,absent4𝛼𝑄\displaystyle=4\alpha\sqrt{Q},= 4 italic_α square-root start_ARG italic_Q end_ARG ,
B0subscript𝐵0\displaystyle B_{0}italic_B start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT =iμ02a0Qis the magnetic field at the center of the coil [G],absent𝑖subscript𝜇02subscript𝑎0𝑄is the magnetic field at the center of the coil [G]\displaystyle=i\mu_{0}\frac{2a_{0}}{\sqrt{Q}}\,\text{is the magnetic field at % the center of the coil [G]},= italic_i italic_μ start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT divide start_ARG 2 italic_a start_POSTSUBSCRIPT 0 end_POSTSUBSCRIPT end_ARG start_ARG square-root start_ARG italic_Q end_ARG end_ARG is the magnetic field at the center of the coil [G] ,
i𝑖\displaystyle iitalic_i is the current in the loop wire [A],
K(k)𝐾𝑘\displaystyle K(k)italic_K ( italic_k ) is the complete elliptic integral function of the first kind, and
E(k)𝐸𝑘\displaystyle E(k)italic_E ( italic_k ) is the complete elliptic integral function of the second kind.

Using the formulas presented in Eq. (1) and Eq. (2), we determined the coil dimensions required to produce an axial MF with a maximum chosen strength of 15 mT. Taking into account several factors, including the desired dimension of our sample (we optimized the design for cells hosted within a 35 mm Petri dish), the size of a typical cell incubator, MF homogeneity and our available water system, we chose the coil dimensions as listed in Tab. 1.

The enclosure of the copper coils is 3D-printed using thermoplastic material, which offers the advantages of being lightweight, corrosion-free, and of not experiencing air condensation at lower temperatures, as opposed to metals. A total of 96 turns of copper wires, each with a diameter of 0.71 mm, were wound around the coil holder and we apply thermally conductive glue on the surfaces of the coils. The wound coils are placed within the coil enclosure, and both electrical and water connections are established. A two-part epoxy is employed to seal the electrical connections. To prevent water leaks, we apply silicone sealant and firmly fixed the caps to the coil enclosure before screwing them. Fig. 1 depicts an exploded cross-section view to scale.

Table 1: Geometrical details of the designed coil.
Parameter Symbol Value
Inner diameter of enclosure (mm) Disubscript𝐷𝑖D_{i}italic_D start_POSTSUBSCRIPT italic_i end_POSTSUBSCRIPT 54
Outer diameter of enclosure (mm) Dosubscript𝐷𝑜D_{o}italic_D start_POSTSUBSCRIPT italic_o end_POSTSUBSCRIPT 114
Width of enclosure (mm) w𝑤witalic_w 20
Copper wire thickness (mm) t𝑡titalic_t 0.71
Inner diameter of coil holder (mm) dcoilsubscript𝑑𝑐𝑜𝑖𝑙d_{coil}italic_d start_POSTSUBSCRIPT italic_c italic_o italic_i italic_l end_POSTSUBSCRIPT 80
Width of coil holder (mm) wcoilsubscript𝑤𝑐𝑜𝑖𝑙w_{coil}italic_w start_POSTSUBSCRIPT italic_c italic_o italic_i italic_l end_POSTSUBSCRIPT 16
Number of turns in each coil N𝑁Nitalic_N 96
Resistance of coil (ohm\mathrm{\SIUnitSymbolOhm}roman_Ω) R𝑅Ritalic_R 2.74±0.01plus-or-minus2.740.012.74\pm 0.012.74 ± 0.01
Voltage drop in coil (V) ΔVΔ𝑉\Delta Vroman_Δ italic_V 2.75±0.01plus-or-minus2.750.012.75\pm 0.012.75 ± 0.01
Coil inductance (μ𝜇\muitalic_μH) L𝐿Litalic_L 870±15.0plus-or-minus87015.0870\pm 15.0870 ± 15.0
Inner diameter of water tube (mm) disubscript𝑑𝑖d_{i}italic_d start_POSTSUBSCRIPT italic_i end_POSTSUBSCRIPT 4
Outer diameter of water tube (mm) dosubscript𝑑𝑜d_{o}italic_d start_POSTSUBSCRIPT italic_o end_POSTSUBSCRIPT 6
Length of water tube (m) l𝑙litalic_l 4
Refer to caption
Figure 1: Exploded cross-sectional view of a single coil in the designed holder with detailed dimensions. It includes the water and electrical connections (1), the enclosure for the coil made of Delrin material (2), the 3D-printed coil holder (3) and the wound copper coil with 96 turns of copper (4). The coil fits within the enclosure.

Fig. 2, left, features the computer-assisted design of the set of coils, highlighting its various components. Fig. 2, right, shows a photo of the design; note the integration of a servo motor to rotate the coil. Additionally, a thermocouple wire has been incorporated in a holder beneath the sample Petri dish for real-time monitoring of temperature fluctuations. Following the assembly of the coils, we did water-leak tests and measured the produced MF when different currents were applied. A 3-axes magnetometer is used to measure the produced MF.

Detailed instructions for coil construction are freely available on our GitHub repository [48].

II.1 Field simulations and measurements

A Helmholtz configuration is achieved by placing the two coils in our arrangement at a distance of 43.4 mm, which is equal to the radius of each coil [49]. In order to confirm the simulated magnetic flux values and the extent to which the produced MFs are homogeneous, we created a numerical model in COMSOL Multiphysics [50]. Our simulated MF values are shown in Fig. 3, for a current of 1 A.

Refer to caption
Figure 2: Computer-assisted design and photo of the water-cooled, 1D Helmholtz coil set. Left, components of the designed coil, namely: (1) water cooling connections; (2) copper coil contained within the enclosure; (3) holder for 35 mm Petri dish; (4) a pivot point demonstrating the swinging motion of the coils while the Petri dish remains stationary; (5) electrical connections; and (6) temperature sensor employed to accurately monitor the temperature directly below the Petri dish within the cell culture area (resolution of 0.1C). Right, fully assembled coil set.

When a simulated current of 1 A flows through the coils, the MF variation over a 35 mm Petri dish has a maximum axial (radial) divergence of 4.2 (1.7  μ𝜇\muitalic_μT. We performed MF measurements along the axial (Basubscript𝐵aB_{\mathrm{a}}italic_B start_POSTSUBSCRIPT roman_a end_POSTSUBSCRIPT) and radial (Brsubscript𝐵rB_{\mathrm{r}}italic_B start_POSTSUBSCRIPT roman_r end_POSTSUBSCRIPT) directions, and the results closely match our simulations. At a current of I=1𝐼1I=1italic_I = 1 A, we measured Basubscript𝐵aB_{\mathrm{a}}italic_B start_POSTSUBSCRIPT roman_a end_POSTSUBSCRIPT = (1.850±0.015)plus-or-minus1.8500.015\left(1.850\pm 0.015\right)( 1.850 ± 0.015 ) mT and Brsubscript𝐵rB_{\mathrm{r}}italic_B start_POSTSUBSCRIPT roman_r end_POSTSUBSCRIPT = (0.000±0.002)plus-or-minus0.0000.002\left(0.000\pm 0.002\right)( 0.000 ± 0.002 ) mT. The measurement error typically grows with the applied current due to imperfections in the coil winding process.

Refer to caption
Figure 3: COMSOL simulation depicting the uniform magnetic field generated by the designed coil set. This simulation, for a current of 1 A, indicates that the MF within the coil is uniformly distributed, measuring (1.80 ±plus-or-minus\pm± 0.01) mT over the 35 mm Petri dish. This uniform MF primarily results from the axial component, as the symmetric nature of the coil cancels out the radial magnetic components.

If changing the MF frequency is desired, it is crucial to acknowledge the physical limitations of producing alternating MFs. One of these limitations is due to the coils’ inductance, which introduces a time delay in the current buildup process. The coils are measured to have inductance L= 870μ𝐿870𝜇L=\,870\,\muitalic_L = 870 italic_μH, resistance R= 2.74Ω𝑅2.74ΩR=\,2.74\,\Omegaitalic_R = 2.74 roman_Ω and a voltage drop of ΔV=2.75VΔ𝑉2.75V\Delta V=2.75\,\text{V}roman_Δ italic_V = 2.75 V. Although a voltage simulation indicates a rapid increase to 2.75 V within a few microseconds, due to the inductive reactance of the coil, achieving a current of 1 A and returning to zero, experimentally, takes about 130 ms when toggling the coil on and off. It is thus essential to ensure that the switching frequency does not exceed f7similar-to𝑓7f\sim 7italic_f ∼ 7 Hz in this situation; otherwise, the current in the coils will be lower than expected. If there is a requirement to operate at higher frequencies, it becomes necessary to adjust the voltage accordingly to achieve the intended current (in other words, to perform impedance matching).

II.2 Efficient heat dissipation in the coils

We analyzed the temperature increase of the water within a coil subjected to electric currents (see Suppl. Sec. SSI). The power dissipated by the coil contributes to heat transfer to water through convection. Experimentally, there is no observed temperature increase exceeding 0.1 C when the coils operate at a current of 1 A, generating approximately similar-to\sim 1.8 mT. Suppl. Fig. SF1 illustrates the simulated temperature rise in the coils based on current and water flow rates, showcasing the system’s performance under various conditions.

II.3 Controlling moisture on coil surfaces

The dew point, denoting the temperature at which air becomes saturated with moisture and condensation initiates [51], needs to be considered in order to avoid water droplets to form onto the sample dish. Additionally, maintaining a higher dew point temperature can avert complications such as condensation-induced corrosion and potential electrical issues. In Suppl. Sec. SSII, we have analyzed the dew point temperature for the coils placed inside the incubator versus the incubator’s own temperature and humidity. Under typical incubation parameters (namely, temperature 37C and humidity 60%), we thus must set the water cooling temperature inside the coil enclosure to be above 28C to avoid moisture condensation (Suppl. Fig. SF2).

III Weak magnetic fields effects in cytoskeleton dynamics

We deploy the coils to study the effects of weak MFs on the fundamental process of cell cytoskeleton assembly, which is central in distinct stages of genesis and in the resolution of many diseases. We use A7r5, a cell line of vascular smooth muscle cells (VSMCs) that is responsive to mechanical stimulation and that has been shown to regulate tensional homeostasis by coupling local oxidant generation to actin cytoskeleton remodeling [52]. To the best of our knowledge, the question of whether cytoskeleton remodeling in VSMCs can be modulated by MF stimulation had not been previously addressed.

III.1 Weak magnetic fields disturb cytoskeleton reassembly after nocodazole exposure

We hypothesized that a weak MF would affect cytoskeletal organization in conditions in which more intense reorganization takes place—e.g., during development or tissue repair responses. To model these conditions, we exposed the VSMCs to a reversible microtubule (MT)-disrupting agent, nocodazole, and observed the dynamics of cytoskeletal recovery after nocodazole withdrawal.

Fig. 4 depicts the experimental protocol, which starts with the seeding of approximately 10,000 A7r5 VSMCs in each well of a 8-well chamber (0.7 cm2/well). After 24 h, the serum-rich (10%) Dulbecco’s Modified Eagle Medium (DMEM) culture medium is replaced with serum-free DMEM solution within the wells for at least 1 h. Followed serum starvation, cells were treated with nocodazole 10 μ𝜇\muitalic_μM or vehicle (DMSO) during 30 min. To eliminate any residual substances, the cells were washed twice with phosphate buffered saline (PBS) before being supplied with fresh serum-free medium. Subsequently, the cells are subjected to a MF of 480 μ𝜇\muitalic_μT, in a direction perpendicular to the bottom of the 8-well chamber, for a duration of 3 hours. The magnetic field strength and exposure duration were chosen guided solely by existing literature [6]. After completing the protocol, the medium was removed and the cells were fixed in paraformaldehyde 4% (15 min at 37 C). Subsequently, wells were washed with PBS, followed by permeabilization with Triton X-100 (0.1% in PBS) for 10 min at 37 C. Once done, cells were washed with PBS and incubated with blocking solution (normal goat serum 10%) for 60 min at room temperature (RT). After removing the blocking solution, the cells were subjected to staining using anti-mouse-tubulin antibody (1:200) for 12 h at 4C in a wet chamber. The primary tubulin antibody was washed twice with PBS; the secondary antibody for tubulin, anti-mouse 488 nm (1:200), as well as nuclear and F-actin markers DAPI and phalloidin 635 nm, respectively, were incubated during 60 min with protection from light and at RT. Finally, wells were washed twice with PBS and slides were mounted with PBS/glycerol (1:1, v/v).

Refer to caption
Figure 4: Experimental procedure for examining cytoskeleton reassembly after nocodazole exposure.
Refer to caption
Figure 5: Weak magnetic fields disturb cytoskeleton reassembly after nocodazole exposure, and influence actin polymerization. (A) Cells exposed to the weak magnetic field exhibit impaired microtubule recovery, F-actin fibers that are less aligned but higher in number, and a lack of mature focal adhesions. Representative images out of 229 cells, obtained over the course of two identical experiments. Microtubule (α𝛼\alphaitalic_α-tubulin shown in red, leftmost panels) recovery was evaluated: in the absence (‘DMSO’) or after a 30 min disruption with nocodazole (‘Noco’); under the ambient magnetic field (‘geofield’) or during exposure to a vertical 480 μ𝜇\muitalic_μT magnetic field for 3 h (‘MF’). Actin cytoskeleton filaments (F-actin shown in green, middle panels) were altered by the weak magnetic field. Nucleus staining (in blue) is depicted in all panels. The rightmost panels depict merged microtubule, actin and nuclei stainings. (B) The weak magnetic field decreases F-actin alignment during recovery after nocodazole. The F-actin alignment (between 0 and 100%percent\%%) was quantified over 229 whole-cell images. The median alignment level for the nocodazole-disrupted cells under the ambient field (14.5%percent\%%) is similar-to\sim 1.6 that of the nocodazole-disrupted cells under the applied magnetic field (8.7%percent\%%), with a p === 0.0278 \leq 0.05 indicated with a star. (C) F-actin polymerization is slightly increased in the presence of a weak magnetic field. Four independent plate reader-based fluorescence assays were performed in 2N (with N === {{\{{3,3,3,12}}\}}) individual wells, with fluorescence intensity being a proxy for actin polymerization levels. After 1 h of exposure to a vertical 480 μ𝜇\muitalic_μT magnetic field, the exposed cells emit similar-to\sim 8%percent\%% more fluorescence than the cells only exposed to the ambient geofield.

After the nocodazole challenge, MFs were observed to affect the reorganization of both the MT network and the F-actin architecture, as shown in Fig. 5.

The fluorescence images in Fig. 5A are representative of 229 cells obtained over the course of two identical technical replicates; they were all taken with identical exposure parameters. They reveal that the MF alone, at least at this intensity (480 μ𝜇\muitalic_μT), direction (vertical axis) and time of exposure (3 h), does not affect the architectures of MTs and F-actin (third row in the figure; cells similar to the control cells shown in the first row). Moreover, MTs were recovered after nocodazole washout in control cells under the ambient geomagnetic field (second row in the figure), as expected, but not in cells that were exposed to the external MF (fourth row in the figure). That is, after a short challenge with nocodazole, the MF induced tubulin accumulation at the cell periphery in patches disconnected with the core of the MTs. Given the strength of these effects and the primary effects of nocodazole in MTs, we propose that MTs are the focus of MF effects and that the loss of proper MT polymer formation and direction could be affected by the weak MF application. In parallel, application of the MF also affected the pattern of F-actin architecture during recovery after nocodazole: there are more pronounced stress fibers (seen as increased F-actin staining), with decreased fiber alignment. We then investigated in greater detail whether the loss of proper actin polymer formation and direction could also be affected by the weak MF application.

We quantified the degree of F-actin alignment using the Image J [53] plugin OrientationJ [54]. Fig. 5B shows the calculated F-actin alignment level (a parameter varying from 0 to 100%percent\%%) over the 229 whole cells, spanning the four experimental conditions (cells with or without nocodazole challenge; cells under the ambient geomagnetic field, or exposed to the external MF). Values for individual cells are indicated with a horizontal line; the violin shapes represent density curves whose width corresponds to the approximate frequency of data in each region. The violin shapes have been truncated at 0 as this is the minimum actin alignment level. The average value for each condition is indicated with a circle; the vertical bars indicate one standard deviation above and below the average value. The medians (which are of relevance as the data are not normally distributed) are indicated with a cross. Notably, the median alignment level for the nocodazole-disrupted cells under the ambient field (14.5%percent\%%) is similar-to\sim 1.6 that of the nocodazole-disrupted cells under the applied MF (8.7%percent\%%). The data not being normally distributed, we employ the Mann-Whitney statistical test to probe for a difference between those two populations, obtaining a p value of 0.0278; the star represents a p \leq 0.05 statistical significance. A similar plot for each one of the two biological replicates composing the full dataset of 229 cells is found in Suppl. Fig. SF3.

Such effects on F-actin dynamics could be promoted either indirectly or directly. Indirectly, a tensegrity response [55] of the cytoskeleton components resulting from MT disruption could modify F-actin organization in order to dynamically redistribute cellular mechanical stresses. As an alternative model, we cannot exclude a potential direct effect of the MF on F-actin fiber assembly. For that, we examined the effect of MF exposure on assays of actin polymerization in the absence of nocodazole, as detailed below in Subsection III.2.

In summary, the weak MF by itself did not strongly affect cytoskeleton dynamics in cells at rest—as it is patent for the cells exposed to the MF but not to nocodazole in the third row of Fig. 5A. In turn, after the nocodazole challenge, the weak MF interfered with the cells’ capacity to recover their MT assembly, and altered their actin filament profile, notably yielding actin fibers that are less aligned yet higher in number.

III.2 Weak magnetic fields increase actin polymerization

Cells exposed to the weak MF (but in the absence of nocodazole) show a small increase in actin polymerization, as measured by plate reader-based actin polymerization assays and depicted in Fig. 5C. We performed four technical replicates of experiments with 2N (with N === {{\{{3,3,3,12}}\}}) individual wells seeded with cells following the same procedure outlined in steps 1–4 in Fig. 4, using DMSO, not nocodazole, in step 3. N wells have been kept under the ambient geomagnetic field and N wells have been exposed to the MF. In these assays, fluorescence intensity is a proxy for actin polymerization levels. The fluorescence of pyrene actin (360 nm excitation and emission at 420 nm) was measured at baseline (3 min), after exposure to the ambient geomagnetic field or to the MF (480 μ𝜇\muitalic_μT for 20 min, data not shown), and after incubation with polymerization buffer during 1 h in the absence or presence of the same weak MF. Stimulation with the MF in the absence of a polymerization buffer does not affect the fluorescence level (data not shown). The plotted points represent the increment of fluorescence promoted by the 1 h-incubation with the polymerization buffer, when the baseline fluorescence is subtracted. For each technical replicate, we plot with a circle the average ratio of fluorescence between the MF-exposed cells and the cells kept under ambient conditions; the vertical bars indicate one standard deviation above and below the average value, obtained with compound error propagation. Again using compound error propagation statistics, we obtain as an average for the four technical replicates a fluorescence ratio of 1.08 ±plus-or-minus\pm± 0.07. The data not being normally distributed, we ran a Mann-Whitney statistical test to check if the control and MF-exposed populations are significantly different. Using the fluorescence intensity, normalized to 100%percent\%%, of the four control conditions, versus the control-normalized average values of fluorescence after MF exposure depicted with circles in Fig. 5C, we find a p value of 0.0211 \leq 0.05.

The fluorescence images in Fig. 5A; the fluorescence images of the other 225 cells used to produce Fig. 5B; and the code used to produce Figs. 5B and C can be found at [48].

IV Conclusion

We designed a water-cooled and incubator-compatible Helmholtz coil setup specifically to enable the study of magnetic fields effects on biological systems. The setup, whose full mechanical and electronic blueprints are freely available, allows for multi-directional exposure of cells to 1D magnetic fields by automated rotation; in turn, the water-cooling mechanism ensures temperature stability, thus eliminating potential artifacts in experiments.

The designed coils setup is used to show the effect of short-term exposure (1 or 3 h) to a weak magnetic field of 480 μ𝜇\muitalic_μT along the vertical direction on the cytoskeleton of a rat vascular smooth muscle cell line. We observe that such a magnetic field influences how microtubules and actin reassemble after depolymerization by nocodazole exposure; and increases actin polymerization, even in the absence of nocodazole, by a small but significant amount.

V Acknowledgments

We thank Paul Weiss for his valuable feedback on the manuscript. This work was supported by: Young Investigator-FAPESP (Fundação de Amparo à Pesquisa do Estado de São Paulo) grant 2018/07230-5, scholarship grants 2020/04280-1, 2022/16629-4, and 2023/05600-8 to L.Y.T.; CEPID (Centro de Pesquisa Inovação e Difusão)/FAPESP grant 2013/07937–8 to F.R.M.L.; Pioneer Science–Instituto D’Or grant, ONR grant N00014-21-1-4011, NSF-Nano EAGER grant 2041158, and NSF-Bio EAGER grant 2114144 to C.D.A.

References

  • Barnothy [1964] M. F. Barnothy, Biological effects of magnetic fields (Springer, 1964).
  • Funk et al. [2009] R. H. Funk, T. Monsees, and N. Özkucur, Electromagnetic effects—from cell biology to medicine, Progress in histochemistry and cytochemistry 43, 177 (2009).
  • Adey [1993] W. R. Adey, Biological effects of electromagnetic fields, Journal of cellular biochemistry 51, 410 (1993).
  • Lednev [1991] V. Lednev, Possible mechanism for the influence of weak magnetic fields on biological systems, Bioelectromagnetics 12, 71 (1991).
  • Albuquerque et al. [2016] W. W. C. Albuquerque, R. M. P. B. Costa, T. d. S. e Fernandes, and A. L. F. Porto, Evidences of the static magnetic field influence on cellular systems, Progress in Biophysics and Molecular Biology 121, 16 (2016).
  • Binhi and Prato [2017] V. N. Binhi and F. S. Prato, Biological effects of the hypomagnetic field: An analytical review of experiments and theories, PloS One 12, e0179340 (2017).
  • Driessen et al. [2020] S. Driessen, L. Bodewein, D. Dechent, D. Graefrath, K. Schmiedchen, D. Stunder, T. Kraus, and A.-K. Petri, Biological and health-related effects of weak static magnetic fields (\leq 1 mTmT\mathrm{mT}roman_mT) in humans and vertebrates: A systematic review, PLoS One 15, e0230038 (2020).
  • Hadi and Simon [2022] Z.-H. Hadi and C. Simon, Magnetic field effects in biology from the perspective of the radical pair mechanism, J. R. Soc. Interface. 19, 20220325 (2022).
  • Zheng et al. [2018] L. Zheng, L. Zhang, L. Chen, J. Jiang, X. Zhou, M. Wang, and Y. Fan, Static magnetic field regulates proliferation, migration, differentiation, and yap/taz activation of human dental pulp stem cells, Journal of Tissue Engineering and Regenerative Medicine 12, 2029 (2018).
  • Bertagna et al. [2021] F. Bertagna, R. Lewis, S. R. P. Silva, J. McFadden, and K. Jeevaratnam, Effects of electromagnetic fields on neuronal ion channels: a systematic review, Ann. N. Y. Acad. Sci. 1499, 82 (2021).
  • Wang and Zhang [2017] H. Wang and X. Zhang, Magnetic fields and reactive oxygen species, Int. J. Mol. Sci. 18, 2175 (2017).
  • Vergallo et al. [2014] C. Vergallo, M. Ahmadi, H. Mobasheri, and L. Dini, Impact of inhomogeneous static magnetic field (31.7–232.0 mTmT\mathrm{mT}roman_mT) exposure on human neuroblastoma SHSY5YSHSY5Y\mathrm{SH-SY5Y}roman_SH - SY5Y cells during cisplatin administration, PLoS One 9, 1 (2014).
  • Gurhan et al. [2021] H. Gurhan, R. Bruzon, S. Kandala, B. Greenebaum, and F. Barnes, Effects induced by a weak static magnetic field of different intensities on HT1080HT1080\mathrm{HT-1080}roman_HT - 1080 fibrosarcoma cells, Bioelectromagnetics 42, 212 (2021).
  • Wang et al. [2008] H.-Y. Wang, X.-B. Zeng, S.-Y. Guo, and Z.-T. Li, Effects of magnetic field on the antioxidant defense system of recirculation-cultured chlorella vulgaris, Bioelectromagnetics 29, 39 (2008).
  • Calabrò et al. [2013] E. Calabrò, S. Condello, M. Currò, N. Ferlazzo, D. Caccamo, S. Magazù, and R. Ientile, Effects of low intensity static magnetic field on FTIRFTIR\mathrm{FTIR}roman_FTIR spectra and ros production in SHSY5YSHSY5Y\mathrm{SH-SY5Y}roman_SH - SY5Y neuronal-like cells, Bioelectromagnetics 34, 618 (2013).
  • Chater et al. [2004] S. Chater, H. Abdelmelek, and K. b. Rhouma, Effects of sub-acute exposure to magnetic field on synthesis of plasma corticosterone and liver metallothionein levels in female rats, Pakistan Journal of Medical Sciences 20, 219 (2004).
  • Teodori et al. [2002] L. Teodori, J. Grabarek, P. Smolewski, L. Ghibelli, A. Bergamaschi, M. De Nicola, and Z. Darzynkiewicz, Exposure of cells to static magnetic field accelerates loss of integrity of plasma membrane during apoptosis, Cytometry: The Journal of the International Society for Analytical Cytology 49, 113 (2002).
  • Fanelli et al. [1999] C. Fanelli, S. Coppola, R. Barone, C. Colussi, G. Gualandi, P. Volpe, and L. Ghibelli, Magnetic fields increase cell survival by inhibiting apoptosis via modulation of Ca2+superscriptCalimit-from2\mathrm{Ca}^{2+}roman_Ca start_POSTSUPERSCRIPT 2 + end_POSTSUPERSCRIPT influx, The FASEB journal 13, 95 (1999).
  • Tofani et al. [2001] S. Tofani, D. Barone, M. Cintorino, M. M. de Santi, A. Ferrara, R. Orlassino, P. Ossola, F. Peroglio, K. Rolfo, and F. Ronchetto, Static and elf magnetic fields induce tumor growth inhibition and apoptosis, Bioelectromagnetics 22, 419 (2001).
  • Lin et al. [2014] T. Lin, L. Wan, X. Qi, W. Shi, and J. Lin, A moderate static magnetic field enhances trail-induced apoptosis by the inhibition of cdc2 and subsequent downregulation of survivin in human breast carcinoma cells, Bioelectromagnetics 35, 337 (2014).
  • Pinton et al. [2008] P. Pinton, C. Giorgi, R. Siviero, E. Zecchini, and R. Rizzuto, Calcium and apoptosis: ERER\mathrm{ER}roman_ER-mitochondria Ca2+superscriptCalimit-from2\mathrm{Ca}^{2+}roman_Ca start_POSTSUPERSCRIPT 2 + end_POSTSUPERSCRIPT transfer in the control of apoptosis, Oncogene 27, 6407 (2008).
  • Yamashita and Saito [2001] K. Yamashita and M. Saito, Effects of middle-level static magnetic field on metabolic activity of mitochondria, Electrical Engineering in Japan 137, 36 (2001).
  • Van Huizen et al. [2019] A. V. Van Huizen, J. M. Morton, L. J. Kinsey, D. G. Von Kannon, M. A. Saad, T. R. Birkholz, J. M. Czajka, J. Cyrus, F. S. Barnes, and W. S. Beane, Weak magnetic fields alter stem cell-mediated growth, Science Advances 5, eaau7201 (2019).
  • Zhang et al. [2014] J. Zhang, C. Ding, L. Ren, Y. Zhou, and P. Shang, The effects of static magnetic fields on bone, Progress in Biophysics and Molecular Biology 114, 146 (2014).
  • Chionna et al. [2005] A. Chionna, B. Tenuzzo, E. Panzarini, M. B. Dwikat, L. Abbro, and L. Dini, Time dependent modifications of hep g2 cells during exposure to static magnetic fields, Bioelectromagnetics 26, 275 (2005).
  • Dini and Abbro [2005] L. Dini and L. Abbro, Bioeffects of moderate-intensity static magnetic fields on cell cultures, Micron 36, 195 (2005).
  • Coletti et al. [2007] D. Coletti, L. Teodori, M. C. Albertini, M. Rocchi, A. Pristerà, M. Fini, M. Molinaro, and S. Adamo, Static magnetic fields enhance skeletal muscle differentiation in vitro by improving myoblast alignment, Cytometry Part A 71A, 846 (2007).
  • Pacini et al. [1999] S. Pacini, G. B. Vannelli, T. Barni, M. Ruggiero, I. Sardi, P. Pacini, and M. Gulisano, Effect of 0.2 T static magnetic field on human neurons: remodeling and inhibition of signal transduction without genome instability, Neuroscience Letters 267, 185 (1999).
  • Surma et al. [2014] S. V. Surma, G. B. Belostotskaya, B. F. Shchegolev, and V. E. Stefanov, Effect of weak static magnetic fields on the development of cultured skeletal muscle cells, Bioelectromagnetics 35, 537 (2014).
  • Taniguchi et al. [2004] N. Taniguchi, S. Kanai, M. Kawamoto, H. Endo, and H. Higashino, Study on application of static magnetic field for adjuvant arthritis rats, Evidence-Based Complementary and Alternative Medicine 1, 187 (2004).
  • Markov and Pilla [1997] M. Markov and A. Pilla, Weak static magnetic field modulation of myosin phosphorylation in a cell-free preparation: calcium dependence, Bioelectrochemistry and Bioenergetics 43, 233 (1997).
  • Jing et al. [2010] D. Jing, G. Shen, J. Cai, F. Li, J. Huang, Y. Wang, Q. Xu, C. Tang, and E. Luo, Effects of 180 mTmT\mathrm{mT}roman_mT static magnetic fields on diabetic wound healing in rats, Bioelectromagnetics 31, 640 (2010).
  • Tassone et al. [2023] E. E. Tassone, J. C. Page, and M. J. Slepian, Assessing the effects of pulsed electromagnetic therapy on painful diabetic distal symmetric peripheral neuropathy: A double-blind randomized controlled trial, Journal of Diabetes Science and Technology 0, 2175 (2023).
  • Feychting et al. [2003] M. Feychting, F. Jonsson, N. L. Pedersen, and A. Ahlbom, Occupational magnetic field exposure and neurodegenerative disease, Epidemiology , 413 (2003).
  • Vergara et al. [2013] X. Vergara, L. Kheifets, S. Greenland, S. Oksuzyan, Y.-S. Cho, and G. Mezei, Occupational exposure to extremely low-frequency magnetic fields and neurodegenerative disease, Journal of occupational and environmental medicine 55, 135 (2013).
  • Leal et al. [2023] T. Leal, C. Langer, and the LUNAR collaboration, Tumor treating fields therapy with standard systemic therapy versus standard systemic therapy alone in metastatic non-small-cell lung cancer following progression on or after platinum-based therapy (LUNAR): a randomised, open-label, pivotal phase 3 study, Lancet Oncol. 24, 1002 (2023).
  • Tai et al. [2021] Y. K. Tai, A. Franco-Obregon, and the OncoFTX collaboration, Modulated TRPC1TRPC1\mathrm{TRPC1}TRPC1 expression predicts sensitivity of breast cancer to doxorubicin and magnetic field therapy: segue towards a precision medicine approach, Front. Oncol. 11, 783803 (2021).
  • Jordan et al. [1999] A. Jordan, R. Scholz, P. Wust, H. Fähling, and R. Felix, Magnetic fluid hyperthermia (MFHMFH\mathrm{MFH}roman_MFH): Cancer treatment with ac magnetic field induced excitation of biocompatible superparamagnetic nanoparticles, Journal of Magnetism and Magnetic materials 201, 413 (1999).
  • Hardiansyah et al. [2014] A. Hardiansyah, L.-Y. Huang, M.-C. Yang, T.-Y. Liu, S.-C. Tsai, C.-Y. Yang, C.-Y. Kuo, T.-Y. Chan, H.-M. Zou, W.-N. Lian, et al., Magnetic liposomes for colorectal cancer cells therapy by high-frequency magnetic field treatment, Nanoscale research letters 9, 1 (2014).
  • Serdyukov and Novitskii [2013] Y. A. Serdyukov and Y. I. Novitskii, Impact of weak permanent magnetic field on antioxidant enzyme activities in radish seedlings, Russian journal of plant physiology 60, 69 (2013).
  • Man et al. [1999] D. Man, B. Man, and H. Plosker, The influence of permanent magnetic field therapy on wound healing in suction lipectomy patients: a double-blind study, Plastic and reconstructive surgery 104, 2261 (1999).
  • Hashimoto et al. [2007] Y. Hashimoto, M. Kawasumi, and M. Saito, Effect of static magnetic field on cell migration, Electrical Engineering in Japan 160, 46 (2007).
  • Fujita et al. [2007] A. Fujita, I. Hirota, Y. Kawahara, and H. Omori, Development and evaluation of intermediate frequency magnetic field exposure system for studies of in vitro biological effects, Bioelectromagnetics 28, 538 (2007).
  • Ruiz-Gómez et al. [2004] M. Ruiz-Gómez, M. Prieto-Barcia, E. Ristori-Bogajo, and M. Martınez-Morillo, Static and 50 HzHz\mathrm{Hz}roman_Hz magnetic fields of 0.35 and 2.45 mTmT\mathrm{mT}roman_mT have no effect on the growth of saccharomyces cerevisiae, Bioelectrochemistry 64, 151 (2004).
  • Racuciu and Creangia [2005] M. Racuciu and D. Creangia, Biological effects of low frequency electromagnetic field in curcubita pepo, in Proceedings of the Third Moscow International Symposium on Magnetism (2005) pp. 278–282.
  • Fontanet et al. [2019] A. Fontanet, J. Marcos, L. Ribó, V. Massana, and J. Campmany, Design and construction of 3D3D\mathrm{3D}3 roman_D helmholtz coil system to calibrate 3D3D\mathrm{3D}3 roman_D hall probes, in Journal of Physics: Conference Series, Vol. 1350 (IOP Publishing, 2019) p. 012167.
  • Jackson [1999] R. H. Jackson, Off-axis expansion solution of laplace’s equation: Application to accurate and rapid calculation of coil magnetic fields, IEEE Transactions on Electron Devices 46, 1050 (1999).
  • bio [2023] https://github.com/QuBiTUCLA/biocoil (2023), accessed: June 27, 2024.
  • Ruark and Peters [1926] A. E. Ruark and M. F. Peters, Helmholtz coils for producing uniform magnetic fields, JOSA 13, 205 (1926).
  • COMSOL AB [2023] COMSOL AB, AC/DC Module User’s Guide, COMSOL AB, Stockholm, Sweden, v. 6.2 ed. (2023), pp. 75-84.
  • Lawrence [2005] M. G. Lawrence, The relationship between relative humidity and the dewpoint temperature in moist air: A simple conversion and applications, Bulletin of the American Meteorological Society 86, 225 (2005).
  • Tanaka et al. [2019] L. Y. Tanaka, T. L. Araujo, A. I. Rodriguez, M. S. Ferraz, V. B. Pelegati, M. C. Morais, A. M. d. Santos, C. L. Cesar, A. F. Ramos, A. M. Alencar, et al., Peri/epicellular protein disulfide isomerase-a1 acts as an upstream organizer of cytoskeletal mechanoadaptation in vascular smooth muscle cells, American Journal of Physiology-Heart and Circulatory Physiology 316, H566 (2019).
  • Rueden et al. [2017] C. T. Rueden, J. Schindelin, M. C. Hiner, B. E. DeZonia, A. E. Walter, E. T. Arena, and K. W. Eliceiri, Imagej2: Imagej for the next generation of scientific image data, BMC Bioinformatics 18, 529 (2017).
  • Püspöki et al. [2016] Z. Püspöki, M. Storath, D. Sage, and M. Unser, Transforms and operators for directional bioimage analysis: A survey, in Advances in Anatomy, Embryology and Cell Biology (Springer International Publishing, 2016) p. 69–93.
  • Ingber [2003] D. E. Ingber, Tensegrity i. cell structure and hierarchical systems biology, Journal of cell science 116, 1157 (2003).
  • Bergman [2011] T. L. Bergman, Fundamentals of heat and mass transfer (John Wiley & Sons, 2011).
  • Mancin et al. [2010] S. Mancin, C. Zilio, A. Cavallini, and L. Rossetto, Heat transfer during air flow in aluminum foams, International Journal of Heat and Mass Transfer 53, 4976 (2010).
  • Andrade et al. [2016] P. A. Andrade, G. Wistbacka, H. Larsson, M. Södersten, B. Hammarberg, S. Simberg, J. G. Švec, and S. Granqvist, The flow and pressure relationships in different tubes commonly used for semi-occluded vocal tract exercises, Journal of Voice 30, 36 (2016).
  • Alduchov and Eskridge [1996] O. A. Alduchov and R. E. Eskridge, Improved magnus form approximation of saturation vapor pressure, Journal of Applied Meteorology and Climatology 35, 601 (1996).

Supplementary materials

SSI Thermal control in coil design

The rise in water temperature in a coil can be calculated using the power dissipated by it. Electric currents generate heat due to resistance in the wire, which is then transferred to the water by convection. The heat produced by the coil depends on both current and resistance. Effective transfer relies on water flow and heat capacity. The power dissipated in each coil equals the rate of internal energy increase added to the rate of energy transfer from the coil to its surroundings through its surface. In other words,

Pdiss=dUdt+S𝑱𝑬𝑑A.subscript𝑃diss𝑑𝑈𝑑𝑡subscriptcontour-integral𝑆𝑱𝑬differential-d𝐴P_{\textrm{diss}}=\frac{dU}{dt}+\oint_{S}\bm{\mathit{J}}\cdot\bm{\mathit{E}}\,dA.italic_P start_POSTSUBSCRIPT diss end_POSTSUBSCRIPT = divide start_ARG italic_d italic_U end_ARG start_ARG italic_d italic_t end_ARG + ∮ start_POSTSUBSCRIPT italic_S end_POSTSUBSCRIPT bold_italic_J ⋅ bold_italic_E italic_d italic_A . (SE1)

The expression Pdisssubscript𝑃dissP_{\textrm{diss}}italic_P start_POSTSUBSCRIPT diss end_POSTSUBSCRIPT represents the power dissipated [W], dU/dt𝑑𝑈𝑑𝑡dU/dtitalic_d italic_U / italic_d italic_t is the rate of increase of internal energy [J/s], S𝑱𝑬𝑑Asubscriptcontour-integral𝑆𝑱𝑬differential-d𝐴\oint_{S}\bm{\mathit{J}}\cdot\bm{\mathit{E}}\,dA∮ start_POSTSUBSCRIPT italic_S end_POSTSUBSCRIPT bold_italic_J ⋅ bold_italic_E italic_d italic_A denotes the rate of energy transfer from the coil to its surroundings [J/s], 𝑱𝑱\bm{\mathit{J}}bold_italic_J is the current density vector [A/m2], 𝑬𝑬\bm{\mathit{E}}bold_italic_E is the electric field vector [V/m], and dA𝑑𝐴dAitalic_d italic_A represents an infinitesimal area element on the surface S𝑆Sitalic_S [m2]. Once the system reaches a stationary state [56], only the second term in Eq. SE1 persists, determining the maximum temperature increase of the coil:

ΔTmax=(IN)2ρ4hfcA3/2.Δsubscript𝑇maxsuperscript𝐼𝑁2𝜌4subscript𝑓𝑐superscript𝐴32\Delta T_{\text{max}}=\frac{(IN)^{2}\rho}{4hf_{c}A^{3/2}}.roman_Δ italic_T start_POSTSUBSCRIPT max end_POSTSUBSCRIPT = divide start_ARG ( italic_I italic_N ) start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT italic_ρ end_ARG start_ARG 4 italic_h italic_f start_POSTSUBSCRIPT italic_c end_POSTSUBSCRIPT italic_A start_POSTSUPERSCRIPT 3 / 2 end_POSTSUPERSCRIPT end_ARG . (SE2)

In the above, I𝐼Iitalic_I denotes the current [A], N𝑁Nitalic_N represents the number of turns in the coil, ρ𝜌\rhoitalic_ρ stands for the electrical resistivity (ΩmΩm\Omega\cdot\text{m}roman_Ω ⋅ m), hhitalic_h is a heat transfer constant ranging from 525W/(m2K)5-25\,\text{W/(m}^{2}\cdot\text{K})5 - 25 W/(m start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT ⋅ K ) [57], fcsubscript𝑓𝑐f_{c}italic_f start_POSTSUBSCRIPT italic_c end_POSTSUBSCRIPT signifies the fill factor, and A𝐴Aitalic_A represents the total cross-sectional area of the coil [m2superscriptm2\text{m}^{2}m start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT]. The fill factor, representing the fraction of the coil’s cross-sectional area filled with conductor material, is given by the formula fc=Nπt2/4Asubscript𝑓𝑐𝑁𝜋superscript𝑡24𝐴f_{c}=N\pi t^{2}/4Aitalic_f start_POSTSUBSCRIPT italic_c end_POSTSUBSCRIPT = italic_N italic_π italic_t start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT / 4 italic_A, where t𝑡titalic_t is the diameter of the wire. The fill factor gauges the efficiency of the conductor material in using the space within the coil. In our coil, the fill factor is 0.74, which falls within the typical range of 0.6 to 0.8. Operating the coils with a current of 1 A to generate a magnetic field of B=1.8𝐵1.8B=1.8\,italic_B = 1.8mT results in a dissipated power of Pdiss=subscript𝑃dissabsentP_{\text{diss}}=\,italic_P start_POSTSUBSCRIPT diss end_POSTSUBSCRIPT =2.74 W. At this level of dissipated power, a flow rate of 1 L/min yields a temperature rise of only 0.4superscript0.40.4\,^{\circ}0.4 start_POSTSUPERSCRIPT ∘ end_POSTSUPERSCRIPTC in the water exiting the coil enclosure (Fig. SF1). As this heat is also radiated into the surrounding air (Eq. SE1), using a thermometer with a resolution of 0.1superscript0.10.1\,^{\circ}0.1 start_POSTSUPERSCRIPT ∘ end_POSTSUPERSCRIPTC, in practice we could not measure any temperature increase in the space between the coils, where the Petri dish is placed during experiments.

Refer to caption
Figure SF1: Simulated temperature rise within coils whose initial temperature is 20superscript2020\,^{\circ}20 start_POSTSUPERSCRIPT ∘ end_POSTSUPERSCRIPTC is in agreement with measurements. Left, the simulated relationship between pressure drop and required flow rate for the 4 m long water pipes (inner cross section of 16  mm2) connected to the coils. The used chiller pump needs to be capable of managing such a pressure drop. Right, numerical simulation of how the coils’ temperature responds to varying levels of applied current up to 5 A, as a function of the water flow rate in the system. For example, at an operational current of 3 A (generating a similar-to\sim 3 ×\times× 1.8 similar-to\sim 5.4 mT magnetic field), a flow rate of 2 L/min effectively keeps the temperature increase within a range of 0.4superscript0.40.4\,^{\circ}0.4 start_POSTSUPERSCRIPT ∘ end_POSTSUPERSCRIPTC for the water exiting the coil. At an operational current of 1 A, no temperature rise of 0.1superscript0.10.1\,^{\circ}0.1 start_POSTSUPERSCRIPT ∘ end_POSTSUPERSCRIPTC or above could be measured at the sample position.

To avoid system overheating, a temperature control sensor has been integrated and attached to the surface of the coil enclosure. The sensor is a passive thermostat (KSD9700), engineered to interrupt the circuit when the coil overheats. With a flow rate set at 1 L/min, the pressure drop across the entire coil configuration, starting from the point where water enters a 4 mm inner diameter pipe and continuing until it exits the enclosure, is simulated to be ΔP0.33barsimilar-toΔ𝑃0.33bar\Delta P\sim 0.33\,\text{bar}roman_Δ italic_P ∼ 0.33 bar. It must be noted that we considered a open flow setup and didn’t take the back pressure [58] that your connected coil might feed into the system.

SSII Addressing condensation on the coils

When using water to cool the coils, it is imperative to prevent air condensation on the enclosure. We determined the lowest temperature (dew temperature) [51] under various incubator humidity levels below which air condensation occurs. Eq. SE3 estimates the dew point temperature (Tdewsubscript𝑇𝑑𝑒𝑤T_{dew}italic_T start_POSTSUBSCRIPT italic_d italic_e italic_w end_POSTSUBSCRIPT) from relative humidity (RH𝑅𝐻RHitalic_R italic_H) and temperature (T𝑇Titalic_T) using empirical constants.

Tdew(°C)subscript𝑇dew(°C)\displaystyle T_{\text{dew}}\,\text{(\textdegree C)}italic_T start_POSTSUBSCRIPT dew end_POSTSUBSCRIPT (°C) =c(log(RH100)+bTc+T)×(blog(RH100)bTc+T)1.absent𝑐𝑅𝐻100𝑏𝑇𝑐𝑇superscript𝑏𝑅𝐻100𝑏𝑇𝑐𝑇1\displaystyle=c\cdot\left(\log\left(\frac{RH}{100}\right)+\frac{b\cdot T}{c+T}% \right)\times\left(b-\log\left(\frac{RH}{100}\right)-\frac{b\cdot T}{c+T}% \right)^{-1}.= italic_c ⋅ ( roman_log ( divide start_ARG italic_R italic_H end_ARG start_ARG 100 end_ARG ) + divide start_ARG italic_b ⋅ italic_T end_ARG start_ARG italic_c + italic_T end_ARG ) × ( italic_b - roman_log ( divide start_ARG italic_R italic_H end_ARG start_ARG 100 end_ARG ) - divide start_ARG italic_b ⋅ italic_T end_ARG start_ARG italic_c + italic_T end_ARG ) start_POSTSUPERSCRIPT - 1 end_POSTSUPERSCRIPT . (SE3)

The numerical constants, a=6.112𝑎6.112a=6.112italic_a = 6.112, b=17.62𝑏17.62b=17.62italic_b = 17.62, and c=243.12𝑐243.12c=243.12italic_c = 243.12, used in this context are sourced from [59]. Our simulations suggest that, in an incubator accurately adjusted to the standard conditions of 37 C and 60 % humidity, the temperature of the water circulating within the coils needs to be above 28 °C in order to prevent the air condensation (Fig. SF2).

Refer to caption
Figure SF2: Estimating the air’s dew point temperature in correlation to the temperature and humidity levels within an incubator. To prevent moist condensation on the surface of the enclosure placed inside an incubator, it is necessary to maintain the water temperature circulating within the enclosure above 28 C. The dashed line illustrates how to adjust the water cooling temperature in the coil for various incubator temperatures while maintaining a constant incubator humidity level at 60%.

SSIII Biological replicates composing the dataset of Fig. 5B

The F-actin alignment data of Fig. 5B was obtained by analysis of 229 cells studied over the course of two biological replicates. We present below a similar plot for each one of the identical experiments.

Refer to caption
Figure SF3: Separate data for each one of the biological replicates composing the dataset of Fig. 5B.