Introduction

The detection rate of prostate cancer (PCa) in men with elevated and/or rising prostate specific antigen (PSA) after negative transrectal ultrasound -guided biopsy (TRUS-Bx) sessions is poor. Hambrock et al. found a cancer detection rate of 59%, in men with an elevated PSA and multiple negative TRUS-Bx (≥2) sessions, for magnetic-resonance guided biopsies (MRGBx). This is an improvement when compared to 8 to 12-core TRUS-Bx schemes with a detection rate around 17% (TRUS-Bx ≥ 1) [13]. Nevertheless, MRGBx is unpleasant for the patient and time-consuming for the radiologist. For these reasons an in-house pneumatically actuated MR-compatible robot was developed where needle-guide direction can be controlled in real-time inside the controller room [4]. Consequently, the patient remains inside the scanner bore. This may decrease procedure time, enhance patient comfort and improve needle-guide positioning.

Eighty-percent of the tumors with a volume larger than 0.5 cm3 (diameter = 1.0 cm) are likely to be clinically significant [5]. Therefore, it is desirable to have a technique with a biopsy error smaller than 5 mm. Different factors, such as needle-guide positioning, patient and prostate motion, and tissue deformation influence the accuracy of needle positioning [68]. Consequently, the needle does not always reach the cancer suspicious region (CSR).

In a phantom study the new robotic technique demonstrated a short manipulation time of 5 min (range 3–8 min) and a high accuracy of 3.0 mm (range 0–5.6 mm) for needle positioning [4]. Yakar et al. demonstrated that it is technical feasible to perform transrectal prostate biopsies using the novel robotic technique (n = 10) (Fig. 1) [9]. To evaluate and optimize the biopsy procedure in the future, it is necessary to identify and quantify the cause of the biopsy error. Therefore, the purpose of this study was to evaluate the accuracy and speed of the novel robotic technique as an aid to perform MRGBx on patients with CSRs.

Fig. 1
figure 1

a the robot with (1) the needle guide, (2) safety mechanism with the suction cup, (3) tapping mechanism to introduce the needle guide, (4) pneumatic motor, (5) tubings to the motors, (6) ground plate for installation on the MR table, (7) angulation rail to move the needle guide in the coronal plane. b set-up of a patient with the robotic technique on the table of the MR system. The patient was positioned in prone position in the MR system. After the needle guide was inserted rectally it was attached to the robot

Materials and methods

Patients

This study was approved by the ethics review board and written informed consent was obtained from all patients who were biopsied with the robotic technique. From February to September 2010, 13 consecutive patients with an elevated PSA (>4 ng/mL) and at least one negative TRUS-Bx session were included. Patient were included in the robotic patient population based on their willingness. The manual patient population was matched to the robotic population. Prior to the MRGBx, patients received a 3 T (Magnetom TRIO, Siemens, Germany) multi-parametric MRI examination of the prostate for identification of possible CSRs. T2-weighted (T2-w) images in three orthogonal planes, transversal diffusion weighted images (DWI) and dynamic contrast-enhanced (DCE)-MR images (Table 1) were obtained during and after injection of 15 ml gadopentetate dimeglumine with a power injector (Guerbet, Gorinchem, Netherlands). To evaluate the CSR for clinical significance, the criteria for MRGBx reported by Hambrock et al. were applied [3].

Table 1 Imaging protocol with sequence specifications. Volumetric images were utilized to identify anatomical landmarks used to quantify target displacement. * The initial two volumetric images (2/13) were gradient echo sequences (first 2 procedures using the robotic technique)

CSRs were determined in consensus by 2 readers with at least 6 years of experience in prostate MR reading. From each CSR the volume (in cc) was determined on the MR images, assuming that the lesions were ellipsoids.

MRGBx

MRGBx were performed within 12 weeks after the diagnostic multi-parametric MR examination. Antibiotic prophylaxis was given with 500 mg ciprofloxacin in the morning and evening for three consecutive days, starting the day before biopsy. A schematic representation of the steps taken to perform a biopsy is illustrated in the flow chart (Fig. 2).

Fig. 2
figure 2

Flow chart of the biopsy procedure for both the robotic and manual techniques. Scan plane adjustments were only perfomed with the IFE software (step 4) when using the robotic technique

A needle-guide filled with gadolinium-doped water was inserted in the rectum of the patient. Subsequently, the needle-guide was mounted to the robotic or manual device(Step 1). The MR imaging protocol (Step 2) for target selection and to navigate on during the biopsy procedure is shown in Table 1.

Targets were selected (Step 3) on these images based on the CSRs found in the diagnostic MRI examination. Manipulation of the needle-guide was done using either the robotic or manual technique (Step 5). After correct alignment of the needle-guide the insertion depth of the needle was measured on a transversal true-FISP (TRUFI) image (Table 1). The patient was slid out the gantry to insert the biopsy needle manually (titanium 18-gauge, fully automatic, core-needle, double-shot biopsy gun with needle length of 175, sampling length of 17 mm (Invivo, Schwerin, Germany).

A T2-w 3D volumetric gradient-echo (first two procedures with the robotic technique) or a T2-w 3D volumetric spin-echo image (all other patients) was acquired with the needle inserted (Step 7). This was the same 3D volumetric image (Table 1) as acquired in step 2. When the needle was in correct position, confirmed on the control images, another target could be targeted (Steps 3–8) or the patient was removed from the MR table (Step 9).

Needle-guide positioning: robotic technique

The robotic system has five degrees of freedom: translations in three directions (anterior-posterior, inferior-superior, lateral) and rotations in two directions (inferior-superior, lateral). The angle of the needle-guide with the main magnetic field could range from 30° to 55° in the inferior-superior direction and plus or minus 26° in the lateral direction [4].

With a simple graphical user interface the direction of the needle-guide can be adjusted [9]. A software package (Interactive Front End (IFE); Siemens Corporate Research, Baltimore (MD), USA) was used to orient and direct the needle-guide in the desired direction (Step 5) under real-time image guidance (Table 1). Manipulation of images and relevant controls can be performed during imaging which allows interactive slice positioning for path planning and real-time monitoring [10, 11].

Needle-guide positioning: manual technique

Transversal and sagittal TRUFI images (Table 1) through the needle-guide were acquired to determine needle-guide direction. The patient was withdrawn from the scanner bore and the radiologist manually adjusted the biopsy device to point the needle-guide towards the target. To confirm correct positioning of the needle-guide sagittal and transversal TRUFI images were acquired through the needle-guide again. These actions were repeated until the needle-guide was in correct position [12].

Total procedure time and manipulation time were recorded for both the robotic and manual techniques.

Measurements

Motion and deformation of the prostate may occur during the biopsy procedure. This will have effect on the position of the target. Since targeting of the CSR in both methods was done on the images acquired in Step 2, which do not take deformation and motion into account, it is important to distinguish between targeting and biopsy error (Fig. 3).

Fig. 3
figure 3

Representation of the needle inside the prostate, illustrating targeting error (ε), target displacement (φ) and biopsy error (δ). The targeting error, defined as the normal distance from needle to the original target coordinate (T), is shown. Target displacement, defined as the distance between original target (T) and transformed target (T′), is represented by φ. Furthermore, the biopsy error (δ) is shown, which is defined as the normal distance between transformed target (T′) and needle

Targeting error (ε)

The targeting error is defined as the normal (shortest) distance from the needle trajectory to the original target location (Fig. 3). This error does not take tissue deformation and patient motion into consideration. This error is a measure for needle-guide positioning towards the intended target.

Biopsy error (δ)

The biopsy error is defined as the normal distance from the needle to the transformed target location (the actual target location after needle insertion). The coordinates of the transformed target are corrected for tissue deformation, as well as patient and prostate motion. The transformed target coordinates were calculated by adding the target displacement vector to the original target coordinates.

Target displacement (φ)

The target displacement vector is defined by the distance and angle between the original and transformed target location. The target displacement (φ) is the length of this vector. The 3D volumetric images made before (Step 2) and after needle insertion (Step 7) were used to determine the target displacement vector. In these images identical anatomical landmarks around the target (mean 13.7 mm; range 2.3 – 48.4 mm) were manually selected with the aid of an open source fusion package [13]. Calcifications, benign prostate hyperplasia (BPH) nodules, the verumontanum and the urethra were used as anatomical landmarks. Coordinates of these anatomical landmarks (≥5) were used to create a 3D vector field. The arrows represent the direction and distance of displacement of anatomical landmarks (Fig. 4). The mean vector of this vector field is a quantitative measure for localized target displacement, since anatomical landmarks around the target were selected.

Fig. 4
figure 4

3D vector field: The blue arrows represent the direction and displacement of the anatomical landmarks. The red arrow is the mean vector representing target displacement. Furthermore, the needle trajectory (black line), targeting error (ε), original target (T), biopsy error (δ) and transformed target (T′) are shown

To calculate the targeting and biopsy error the needle trajectory was determined by fitting a line through multiple points (≥8) within the needle artifact using linear regression in 3D-space. These points were obtained from the 3D volumetric MR images obtained in Step 7.

Target coordinates (CSRs) were obtained from the T2-w and DW images acquired in Step 2.

The angle between needle trajectory and target displacement direction was calculated in order to see whether the target moved along the needle trajectory or in a random direction (Fig. 5).

Fig. 5
figure 5

Transversal TRUFI image through the needle guide before (a) and after (b) needle insertion in the prostate

Statistical analysis

Two-tailed independent t-tests were performed to determine whether there were significant differences between the robotic and manual techniques for targeting error, biopsy error, target displacement, procedure and manipulation time. Differences were considered to be significant at p < .05. Statistical analysis were performed with SPSS, version 16.0.01 (Chicago, Illinois).

Results

In total, 13 patients with 32 needle positions were analyzed. Table 2 describes the patient characteristics and biopsy results for both techniques.

Table 2 Patient characteristics and biopsy results for both the robotic and manual techniques

Since the majority of previous negative TRUS-Bx had been performed outside our institution, information on the number of cores was not available for most patients.

Accuracy

The mean targeting error for both the robotic and manual techniques was almost similar (5.7 vs 5.8 mm respectively, p = 0.928) (Fig. 6). The mean biopsy error was less (4.4 vs 6.5 mm) with the manual technique compared to the robotic technique (p = 0.054). Target displacement was larger with the robotic technique (6.6 vs 6.0 mm, p = 0.439).

Fig. 6
figure 6

Histogram showing the mean targeting error, biopsy error and target displacement for both the robotic and manual techniques. The error bars represent the standard deviation

Direction of target displacement

The mean angle between needle trajectory and target displacement direction for the robotic and manual techniques was 36.7° (range 4.0–82.2°) and 37.6° (range 7.7–73.3°), respectively.

Time

The mean time to perform a biopsy procedure using the robotic technique was 76 min (range 60–100 min) and 61 min (range 52–64 min) with the manual technique. The total procedure time includes the extra time to acquire the 3D volumetric images (2:36 min each). The mean manipulation time to move from target to target was 6 min (range 3–11 min) with the robotic technique and 8 min (range 5–11 min) with the manual technique. The differences in manipulation time and procedure time between both techniques were not significant.

Discussion

The robotic and manual techniques demonstrated comparable results regarding targeting error and target displacement. The biopsy error was larger when using the robotic technique, however not statistically significant. The robotic technique prevented the need of moving the patient in and out of the scanner bore for manipulation and imaging of the needle-guide. Most of the target displacement found in our study was in the direction of the needle trajectory.

Several robots for transperineal seed delivery in brachytherapy have been described in literature [1417]. The robotic and manual techniques for transrectal biopsies demonstrated a larger targeting error (5.7 and 5 mm respectively) compared to other robotic techniques. Muntener et al. found a targeting error of 2.02 mm (range 0.86–3.18 mm) with their robot in a canine model [16]. Zangos et al. describe a transgluteal approach for prostate biopsy with a targeting error of 0.9 mm (range, 0.3–1.6 mm) [18]. The targeting error, biopsy error and target displacement for a transrectal biopsy device were 2.2 mm (range 0.5–5.7 mm), 5.1 mm (range 1.6–11.0 mm) and 5.4 mm (range 1.6–11.1 mm) respectively. Although the targeting error was less compared to our results, the biopsy error and target displacement were in concordance with our results [19].

MR-guided TRUS-Bx may be an alternative to MRGBx in the future because its availability and is probably less expensive. However, initial results show registration errors around three millimeter in phantoms and patients [2023]. In addition to this error is the targeting error and tissue deformation which together determine the ability to sample a CSR.

The anatomical landmarks chosen in the MR images to determine target displacement were selected manually. This may have introduced an error, since it is difficult to select exactly the same position. Automatic registration would be an alternative to diminish this error. However, automatic registration is difficult and introduces errors as well [15]. Furthermore, the images that need to be registered are different in the area of the target, because of the presence of the needle, causing a line shape void in the area where best registration is needed.

The biopsy procedures with the robotic and manual techniques were not performed by the same radiologist. To overcome the limitation of inter-variability, the performing physician of each procedure performed the biopsy session in consensus with the first author who attended all sessions.

Despite the fact that patient selection for each technique was not randomly chosen, both groups had similar patient characteristics (Table 2). The number of cores taken during TRUSBx was not available for most patients in both groups.

In the first patient a gradient-echo sequence was used to determine needle trajectory and target displacement. The needle artifact size was acceptable, varying from 3.5 mm to 4.5 mm. In the second patient, the angle of the needle with the static magnetic field was larger. As a result, the artifact size of the needle increased to 8.5 mm. Therefore, we decided to use a 3D spin-echo sequence in which the signal void around the needle was less influenced by distortions of the magnetic field. Needle artifact size now varied from 3.8 to 4.7 mm.

The quantitative method described to determine target displacement cannot discriminate between patient motion, prostate motion and tissue deformation. However, our results demonstrated that most target displacement was in the direction of the needle trajectory suggesting that most of the target displacement was caused by needle insertion.

Hambrock et al. found a median imaging time of 35 min for MRGBx [24]. In our study, we reported the total procedure times (including patient preparation) for the robotic and manual techniques. Although manipulation time was shorter when using the robotic technique, the total procedure time was longer compared to the manual technique. Positioning of the patient was a precise and time-consuming process. In case of incorrect positioning the whole set-up did not fit inside the scanner bore, or the range of motion of the robotic technique was impaired. Furthermore, a connection with the IFE software was necessary for real-time image guidance. Even for small adjustments in needle-guide direction, a interaction with the IFE was necessary including manual selection of the correct slice direction through the needle-guide. Furthermore, some actions were only possible from behind the MR-console (e.g. measurement tool); the operator had to switch constantly between MR-console and IFE monitor during the procedure. Altogether this led to an extension of the procedure time.

Image registration during the biopsy procedure can correct for target displacement and may attribute to reduce the biopsy error [25]. Nevertheless, image registration is often a time-consuming process. Furthermore, our results suggested that movement of the target was mainly caused by needle insertion. Image registration would not correct for prostate motion due to needle insertion. Deformation models of the prostate to predict tissue deformation due to needle insertion may help to overcome this problem [26]. Other alternatives are different techniques for needle insertion, such as rotating needles and tapping devices [2730].

Promising treatment types in the MR-scanner such as focal cryosurgery [31] and laser ablation [32] are now under investigation. Major advantages of treatment in the MR scanner are the ability of soft tissue imaging and monitoring (for example temperature mapping) [33]. Robotics will play an important role in the future during treatment in the MR since accurate needle placement is required.

In spite of the fact that the results are comparable regarding accuracy and speed, the larger biopsy error and the extended technical effort of the robotic technique make the manual technique – currently – more suitable to perform MRGBx. Furthermore, this study provided a better insight in displacement of the target during in-vivo biopsy procedures.