. 2014 Apr 4;9(4):e93757.
doi: 10.1371/journal.pone.0093757. eCollection 2014.

The sensitivity and specificity of using a computer aided diagnosis program for automatically scoring chest X-rays of presumptive TB patients compared with Xpert MTB/RIF in Lusaka Zambia

Affiliations

The sensitivity and specificity of using a computer aided diagnosis program for automatically scoring chest X-rays of presumptive TB patients compared with Xpert MTB/RIF in Lusaka Zambia

Monde Muyoyeta et al. PLoS One. .

Abstract

Objective: To determine the sensitivity and specificity of a Computer Aided Diagnosis (CAD) program for scoring chest x-rays (CXRs) of presumptive tuberculosis (TB) patients compared to Xpert MTB/RIF (Xpert).

Method: Consecutive presumptive TB patients with a cough of any duration were offered digital CXR, and opt out HIV testing. CXRs were electronically scored as normal (CAD score ≤ 60) or abnormal (CAD score > 60) using a CAD program. All patients regardless of CAD score were requested to submit a spot sputum sample for testing with Xpert and a spot and morning sample for testing with LED Fluorescence Microscopy-(FM).

Results: Of 350 patients with evaluable data, 291 (83.1%) had an abnormal CXR score by CAD. The sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of CXR compared to Xpert were 100% (95%CI 96.2-100), 23.2% (95%CI 18.2-28.9), 33.0% (95%CI 27.6-38.7) and 100% (95% 93.9-100), respectively. The area under the receiver operator curve (AUC) for CAD was 0.71 (95%CI 0.66-0.77). CXR abnormality correlated with smear grade (r = 0.30, p<0.0001) and with Xpert CT(r = 0.37, p<0.0001).

Conclusions: To our knowledge this is the first time that a CAD program for TB has been successfully tested in a real world setting. The study shows that the CAD program had high sensitivity but low specificity and PPV. The use of CAD with digital CXR has the potential to increase the use and availability of chest radiography in screening for TB where trained human resources are scarce.

PubMed Disclaimer

Conflict of interest statement

Competing Interests: The authors have declared that no competing interests exist.

Figures

Figure 1
Figure 1. ROC on a test set of 969 CXRs.
Figure 2
Figure 2. Patient flow.
Figure 3
Figure 3. ROC curve comparing CAD score to Xpert TB as the reference standard.

Similar articles

Cited by

References

    1. Iademarco MF, O'Grady J, Lonnroth K (2012) Chest radiography for tuberculosis screening is back on the agenda. Int J Tuberc Lung Dis 16: 1421–1422. - PubMed
    1. Toman K, Frieden T (2004) Toman's tuberculosis: case detection, treatment and monitoring: questions and answers; TFrieden T, editor. Geneva: World Health Organisation.
    1. Jaeger S, Karargyris A, Candemir S, Siegelman J, Folio L, et al. (2013) Automatic screening for tuberculosis in chest radiographs: a survey. Quant Imaging Med Surg 3: 89–99. - PMC - PubMed
    1. Hogeweg L, Mol C, de Jong PA, Dawson R, Ayles H, et al. (2010) Fusion of local and global detection systems to detect tuberculosis in chest radiographs. Med Image Comput Comput Assist Interv 13: 650–657. - PubMed
    1. Jaeger S, Karargyris A, Candemir S, Folio L, Siegelman J, et al... (2013) Automatic Tuberculosis Screening Using Chest Radiographs. IEEE Trans Med Imaging. - PubMed

Publication types

MeSH terms

Grants and funding

Funding sources: T9-370-114 World Health Organization through the stop TB partnership-TB REACH wave 2 grant. http://www.stoptb.org,/global/awards/tbreach/grantees2.asp, Sub grant- Delft Imaging systems-CAD4TBII, http://www.delftimagingsystems.com/tb-solutions/cad4tb. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

LinkOut - more resources