Hardware and software artifacts in storage phosphor radiography
- PMID: 1947316
- DOI: 10.1148/radiographics.11.5.1947316
Hardware and software artifacts in storage phosphor radiography
Abstract
Hardware and software artifacts in digital radiographs acquired with storage phosphor systems can seriously impair image quality and imitate or mask abnormalities. These artifacts are caused by image plate, image reader, and laser printer defects; faulty image readout; processing errors; and unsharp masking. The artifacts can simulate calcifications and pneumothoraces or conceal low-contrast ill-defined lesions and subtle lesions along opacity interfaces. Hardware artifacts need to be recognized and properly traced to repair the system or improve its maintenance. Artifacts due to software characteristics and image post-processing must also be identified to ensure adequate system handling and adjustment of postprocessing algorithms.
Similar articles
-
Computed radiography image artifacts revisited.AJR Am J Roentgenol. 2011 Jan;196(1):W37-47. doi: 10.2214/AJR.10.5563. AJR Am J Roentgenol. 2011. PMID: 21178029 Review.
-
Analysis of photostimulable phosphor plate image artifacts in an oral and maxillofacial radiology department.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008 Nov;106(5):749-56. doi: 10.1016/j.tripleo.2008.01.003. Epub 2008 Jun 11. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008. PMID: 18547836
-
The evolution of digital radiography: from storage phosphors to flat-panel detectors.Rays. 2003 Jan-Mar;28(1):5-12. Rays. 2003. PMID: 14509175
-
[Digital thorax radiography: flat-panel detector or storage phosphor plates].Radiologe. 2003 May;43(5):351-61. doi: 10.1007/s00117-003-0893-8. Radiologe. 2003. PMID: 12764583 German.
-
Quality control of storage phosphor digital radiography systems.J Digit Imaging. 1995 May;8(2):67-74. doi: 10.1007/BF03168129. J Digit Imaging. 1995. PMID: 7612704 Review.
Cited by
-
Analysis of photostimulable phosphor image plate artifacts and their prevalence.World J Clin Cases. 2022 Jan 14;10(2):437-447. doi: 10.12998/wjcc.v10.i2.437. World J Clin Cases. 2022. PMID: 35097068 Free PMC article.
-
Effect of enveloping and disinfection methods on artefact formation on enveloped PSP plate images.Oral Radiol. 2022 Oct;38(4):558-564. doi: 10.1007/s11282-022-00587-1. Epub 2022 Jan 22. Oral Radiol. 2022. PMID: 35064886 Free PMC article.
-
Imaging evaluating of the implant/bone interface-an in vitro radiographic study.Dentomaxillofac Radiol. 2017 Jul;46(5):20160296. doi: 10.1259/dmfr.20160296. Epub 2017 Mar 28. Dentomaxillofac Radiol. 2017. PMID: 28211288 Free PMC article.
-
Definition, classification and retrospective analysis of photostimulable phosphor image artefacts and errors in intraoral dental radiography.Dentomaxillofac Radiol. 2017 Mar;46(3):20160188. doi: 10.1259/dmfr.20160188. Epub 2017 Feb 17. Dentomaxillofac Radiol. 2017. PMID: 27883287 Free PMC article.
-
Artifacts found during quality assurance testing of computed radiography and digital radiography detectors.J Digit Imaging. 2009 Aug;22(4):383-92. doi: 10.1007/s10278-008-9109-0. Epub 2008 Apr 22. J Digit Imaging. 2009. PMID: 18427900 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources