Soft-copy reading of digital chest radiographs: effect of ambient light and automatic optimization of monitor luminance
- PMID: 15714093
- DOI: 10.1097/01.rli.0000153658.15272.91
Soft-copy reading of digital chest radiographs: effect of ambient light and automatic optimization of monitor luminance
Abstract
Objective: We sought to evaluate whether soft-copy reading of simulated pulmonary chest lesions is influenced by ambient light and automatic optimization of cathode ray tube (CRT) monitor luminance.
Materials and methods: Four types of simulated lesions (nodules, lines, micronodules, and patchy opacities) were superimposed over an anthropomorphic chest phantom. Lesion detection with soft-copy reading was assessed using a high-contrast grayscale 2K CRT monitor under the following conditions: (1) subdued lighting (<50 lux); (2) normal lighting conditions (450 lux) without, and (3) with a sensitivity modulation to automatically adjust the CRT luminance to the increased amount of ambient light. Reading data were analyzed according to receiver operating curve. Significance of differences was tested using an analysis of variance for repeated measures.
Results: Ambient room light of 450 lux did not significantly influence the detection of nodules and patchy opacities. However, bright ambient light significantly decreased detection of micronodules (0.60 vs. 0.74) and lines (0.52 vs. 0.66) relative to subdued lighting conditions. Automatic luminance adjustment could compensate the effect of ambient light for the micronodules (0.77) but not for the lines (0.53).
Conclusion: Bright ambient light significantly decreases detection of small low-contrast structures. This may be partially but not completely compensated by an automatic luminance adaptation.
Similar articles
-
A systematic review of viewing conditions and monitor specifications in mammography.Radiography (Lond). 2020 Nov;26(4):325-331. doi: 10.1016/j.radi.2020.03.001. Epub 2020 Apr 15. Radiography (Lond). 2020. PMID: 32305425 Review.
-
Effect of display type and room illuminance in chest radiographs.Eur Radiol. 2016 Sep;26(9):3171-9. doi: 10.1007/s00330-015-4150-0. Epub 2015 Dec 10. Eur Radiol. 2016. PMID: 26662032
-
Comparison of observer performance on soft-copy reading of digital chest radiographs: high resolution liquid-crystal display monitors versus cathode-ray tube monitors.Eur J Radiol. 2008 Apr;66(1):13-8. doi: 10.1016/j.ejrad.2007.05.023. Epub 2007 Aug 6. Eur J Radiol. 2008. PMID: 17689217
-
Effect of monitor luminance and ambient light on observer performance in soft-copy reading of digital chest radiographs.Radiology. 2004 Sep;232(3):762-6. doi: 10.1148/radiol.2323030628. Epub 2004 Jul 23. Radiology. 2004. PMID: 15273338
-
Impact of ambient light and window settings on the detectability of catheters on soft-copy display of chest radiographs at bedside.AJR Am J Roentgenol. 2003 Nov;181(5):1415-21. doi: 10.2214/ajr.181.5.1811415. AJR Am J Roentgenol. 2003. PMID: 14573447
Cited by
-
Visual assessment method of angular performance in medical liquid-crystal displays by use of the ANG test pattern: effect of ambient illuminance and effectiveness of modified scoring.Radiol Phys Technol. 2014 Jan;7(1):51-6. doi: 10.1007/s12194-013-0230-5. Epub 2013 Aug 10. Radiol Phys Technol. 2014. PMID: 23934325
-
Letter to the editor re: the iPad tablet computer for mobile on-call radiology diagnosis? Auditing discrepancy in CT and MRI reporting.J Digit Imaging. 2013 Apr;26(2):141-2. doi: 10.1007/s10278-013-9581-z. J Digit Imaging. 2013. PMID: 23408011 Free PMC article. No abstract available.
-
Breast screen new South wales generally demonstrates good radiologic viewing conditions.J Digit Imaging. 2013 Aug;26(4):759-67. doi: 10.1007/s10278-013-9571-1. J Digit Imaging. 2013. PMID: 23319112 Free PMC article.
-
Human contrast-detail performance with declining contrast.Med Phys. 2012 Sep;39(9):5446-56. doi: 10.1118/1.4742851. Med Phys. 2012. PMID: 22957612 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials